Table 8-2
Ceiling Suspension Motor Controller Boards
CODE
USER ERROR MESSAGE
11100
Control Board Error: CS-Longitudinal
11101
Control Board Error: CS-Transversal
11102
Control Board Error: CS-Vertical
11103
Control Board Error: CS-Angular
11104
Control Board Error: CS-Rotational
Invalid Calibration Data:
11105
CS-Longitudinal
Invalid Calibration Data:
11106
CS-Transversal
11107
Invalid Calibration Data: CS-Vertical
11108
Invalid Calibration Data: CS-Angular
11109
Invalid Calibration Data: CS-Rotational
Operative rejected: CS-Longitudinal
position out of calibrated range:
11110
Right there, only MANUAL
movement allowed
11111
Board failure: CS-Longitudinal
11112
End-switch limit: CS-Longitudinal
0242F EN 20181218
ERROR DESCRIPTION
Motor Controller Board not connected
or I2C communication problems
Motor Controller Board not connected
or I2C communication problems
Motor Controller Board not connected
or I2C communication problems
Motor Controller Board not connected
or I2C communication problems
Motor Controller Board not connected
or I2C communication problems
Axis not calibrated,
calibration file corrupted
Axis not calibrated,
calibration file corrupted
Axis not calibrated,
calibration file corrupted
Axis not calibrated,
calibration file corrupted
Axis not calibrated,
calibration file corrupted
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
DX-D 600
User Manual
ACTION
Check the Motor Controller I/F
Connection
Check the Motor Controller I/F
Connection
Check the Motor Controller I/F
Connection
Check the Motor Controller I/F
Connection
Check the Motor Controller I/F
Connection
Calibrate once again
If message persists change SBC/CF
Calibrate once again
If message persists change SBC/CF
Calibrate once again
If message persists change SBC/CF
Calibrate once again.
If message persists change SBC/CF
Calibrate once again
If message persists change SBC/CF
Check position feedback
Contact Service
Change board
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range
145