AGFA DX-D 600 User Manual page 112

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

DX-D 600
User Manual
Illustration 5-2
Tabletops
Standard Tabletop
104
TABLETOP
The Patient Support can be longitudinally and transversely moved, allowing an
easy patient positioning. It can be also raised up to a maximum height of
920 mm (36.22") and lowered to a minimum height of 580 mm (22.83"). There
are three different tabletop models available:
Standard Tabletop, which is the default option. It is made of a carbon fibre
sheet and two metal frames at the laterals. Its attenuation is <0.65 mm
eq. Al at 100 kV.
Carbon fiber flat Tabletop. It is composed by a sheet of carbon fiber,
without metallic frames at the laterals. Its attenuation is <0.6 mm eq. Al
at 100kV.
Laminated flat Tabletop. It is composed by a sheet of laminated
melamine. Its attenuation is <1.2 mm eq. Al at 100 kV.
CONTROL PEDALS
The Control Pedals are used to release the brakes and let the Tabletop free
movement for its longitudinal and transverse positioning and vertical motion.
There are four different Control Pedals that allow all vertical and horizontal
movements and positioning: Tabletop Motions, Raise and Lower.
Depending on the configuration, it may be necessary to step twice or once the
Control Pedal to release the brake. It is factory configured to step twice, which
is recommendable for safety reasons to avoid occasional brakes release, but
configurable in the installation.
DOUBLE CONTROL PEDALS (OPTION)
Optionally, two sets of Control Pedals can be installed, one at the front and one
at the back of the RAD Table.