6.4
DR DETECTOR ROTATION
.
Note
Illustration 6-6
DR Detector Rotation Lock Lever and Rotation Detent
1
Rotation Lock Lever
Illustration 6-7
Goniometer for DR Detector Rotation
0242F EN 20181218
This function allows to rotate the DR Detector around the center of the image,
the movement is manually controlled. It is possible to have the DR Detector
o
rotated up to 90
.
To avoid degradation of image quality and loss of DR Detector
functionality, it is recommended to perform exposures with the DR
Detector exactly at 0
can be performed in another range position.
1.
Release the Rotation Lock lever. Lower it to release the Rotation Lock.
2.
Unlock the Rotation Detent, located at the right of the DR Detector
Support. It reinforces the DR Detector Rotation Lock performance and
locks the DR Detector at 0
3.
Rotate DR Detector to the desired position. The Rotation angle is
indicated by the goniometer at the back of the DR Detector.
4.
Lift the Rotation Lock lever to lock the DR Detector Position.
o
o
or 90
positions, even though exposures
o
position.
DX-D 600
User Manual