AGFA DX-D 600 User Manual page 158

Hide thumbs Also See for DX-D 600:
Table of Contents

Advertisement

DX-D 600
User Manual
Table 8-4
Table Motor Controller Boards
CODE
USER ERROR MESSAGE
Control Board Error: Table
11300
Invalid Calibration Data:
11301
Table-Longitudinal
11302
Invalid Calibration Data: Table-Vertical
Operative rejected: Table-Longitudinal
position out of calibrated range:
11310
Right there, only MANUAL
movement allowed
Control Board Error:
11311
Table-Longitudinal
11312
End-Switch Limit: Table-Longitudinal
11313
Position Out of Range: Table-Vertical
11314
Control Board Error: Table-Vertical
150
ERROR DESCRIPTION
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
ACTION
Replace Control Board
Check Longitudinal Axis calibration
Check the Table's Vertical Axis
calibration
Check Longitudinal Axis calibration
Replace the Control Driver Board
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range
Check Axis calibration
Replace the Control Driver Board
0242F EN 20181218