Select Measurement Rate

Background
1. Select Rate

Display Mode

Background
1. Toggle Display
mode
The resistance measurement speed has 2 ranges: slow and
fast. Slow speed is the most accurate with 10
measurements/second. Fast speed has 60
measurements/second. Both have the same measurement
resolution.
The rate selection function is not applicable in Diode
measurement mode. When the PWM drive signal is used
or when the Scan function is activated, the only available
rate setting is fast.
Press the
key to toggle between the Slow and Fast
Speed
rates.
The Display key can be used to toggle between the
normal and the simplified display mode. The simplified
display mode clears all text, menus and function
indicators from the screen except for the measurement
and measurement mode indicators.
Press the
Display
normal and simplified. The display will change
accordingly.
Measurement rate
key to toggle the display between
MEASUREMENT