4.2 Start Up Screens

The Analyzer has a built-in self-diagnostic testing routine.
Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and – at the end of the
test the status of each function – is displayed on the screen, either as OK
or as a number between 1 and 3. (See System Self Diagnostic Test in
Chapter 5, Section 5.4 for number code).
The self-diagnostics are run by demand by the operator whenever
the instrument is turned on. It is also run during analyze mode through
the system menu.
The first screen identifies the make and model of Analyzer and has
three touch sensitive buttons along the bottom.
This screen updates the TCD Chamber temperature, and a settling
time runs down from 30 minutes. When the settling time reaches 0,
system will update the settling time as "completed." After 10 seconds,
the front panel moves to next screen.
Teledyne Analytical Instruments
Figure 4-1: Start Up Screen
2000RS
28