• If the vacuum test is not successful, the replacement head was not sealed 
Testing and
The level of UV radiation from the lamps used for curing models can change 
Calibrating the
over time. To ensure optimum curing of models during printing, a pop-up 
UV Lamps
message reminds you to test the lamps and calibrate the level of UV radiation 
after every 300 hours of printing. You do this by running the UV Calibration 
Wizard.
To measure the UV radiation, you need:
• Stratasys TOL-03005-S (UV sensor and cable)
• a stand-alone UV radiation meter, approved for use with the printer
The wizard compares the measured radiation to the recommended radiation 
level for each of the lamps, at each printing mode—High Speed, and High 
Quality and Digital Material.
When calibrating the lamps, the wizard attempts to adjust the radiation level, 
if necessary.
• If the reading is within the acceptable range, the wizard continues to the 
• If further adjustment is necessary, the current phase is repeated. 
• If the level of UV radiation is too low to be properly adjusted, the wizard 
Before testing and calibrating the UV lamps: 
 
Make sure the glass lens on each of the UV lamps is clean. For easy 
l
access, run the Head Cleaning Wizard, and use the special cleaning pad 
supplied.
 
Make sure that the glass over the UV radiation sensor is clean.
l
DOC-07027 Rev. E
properly during installation, and the following warning screen appears.
Figure 6-87    Vacuum-leakage warning screen
If this happens: 
a.  Click Next.
b.  Follow the instructions on the wizard screens to re-install the head.
or—
next phase. 
continues to the next phase, but the final wizard screen indicates that the 
lamp's radiation for the printing mode is unacceptable.
6 Operating and Maintaining the Objet500 3D Printer