Figure 6-97    UV measurement progress
When using a stand-alone radiation meter, reset the meter and enter the 
reading (measurement) when you are prompted to do so.
Figure 6-98    UV Measurement dialog box
If the UV level measured is not acceptable, the wizard calibrates the UV lamp 
by adjusting the current supplied to it, and then tests it again.
When all UV tests have finished, the results are displayed, showing the 
condition of the lamps after calibration (see Figure 6-99 below).
Figure 6-99    Results and condition of UV lamps after calibration
After examining the UV calibration results:
1.  Click Next.
2.  Remove the UV sensor and close the printer cover. After confirming the 
DOC-07027 Rev. E
items listed in the wizard screen, click Next.