Minimum Cable Configuration for System to Boot

The following cables must be connected between the INT I/O board and the rest of the system:
Cables to Hard Disk (inside Back End Processor)
Patient I/O cable to Patient I/O box
I/O RS-232 serial cable to Front Panel
ACCTRL cable between INT I/O and ACCTRL (power-on signal)
DC Power cable to Front Panel
Cables to the monitor
All other cables can be removed and the system will still boot.
7-12-7
Minimum Cable Configuration for System to Scan
The following cables must be connected between the INT I/O board and the rest of the system:
Cables to Hard Disk (inside Back End Processor)
Patient I/O cable to Patient I/O box
I/O RS-232 serial cable to Front Panel
ACCTRL cable between INT I/O and ACCTRL (power-on signal)
DC Power cable to Front Panel
Cables to the monitor
PCI Cable from BEP to SCB