26.2 Configuration

26.2.1 Connection

A. DUT input and output channels
B. Channel gain. When it is set to Auto, the oscilloscope will automatically adapt the
vertical scale according to the signal amplitude. When it is set to Hold, it will always
keep the currently selected vertical scale. If signals overload the selected range in
Hold mode, the data may not accurately represent the physical test.
C. Arbitrary waveform generator connection settings. Touch Interface to select the
connection type. When selecting LAN, it needs to Set IP and Save it. Touch Test to
confirm whether the arbitrary waveform generator is connected correctly.

26.2.2 Sweep

Touch Sweep Type to select the sweep type. There are two types: Simple and variable
level.
A. Set the sweep type to simple
B. Set the sweep frequency. The frequency mode can be linear or decade. When it is
set to linear, the corresponding center frequency and span frequency need to be set.
When it is set to decade, the corresponding start frequency and stop frequency need
to be set