Delete Polygon: Deletes selected polygon
To edit a vertex, side or polygon object, firstly select it, and then move it by a dragging
gesture or by entering the desired value in coordinate edit area. For a side, the value is
for its middle point. For a polygon, the ordinate is for its geometric center.

23.3 Pass/Fail Rule

The Pass/Fail rule is specified at the Type region in the Mask Test dialog box.
All In: All data points must be inside the mask to pass the test. Even a single point
outside the mask will cause a failure.
All Out: All data points must be outside the mask to pass the test. Even a single point
inside the mask will cause a failure.
Any In: Any data point inside the mask will be recognized as a pass. All data points
outside the mask will cause a failure.
Any Out: Any data point outside the mask will be recognized as a pass. All data points
inside the mask will cause a failure.

23.4 Operation

Touch Operation to start/stop the test. Stopping a test in progress and restart the test will
clear the count of the passing frames, failed frames, total frames, and the fail rate.
Pressing the Clear Sweeps button on the front panel can also clear the count information.
1.800.561.8187
Select a vertex (point)
www.
Select a side (line segment)
Select a polygon
.com
T3DSO2000A Manual. Page 146
information@itm.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lecroy t3dso2204aLecroy t3dso2502aLecroy t3dso2354aLecroy t3dso2104aT3dso2104aT3dso2204a ... Show all

Table of Contents