Great than a time value(≥)-- Trigger when the
positive or negative slope time of the input signal is
greater than the specified time value.
Within a range of time value([--,--])-- Trigger when
the positive or negative slope time of the input signal
is greater than the specified lower limit of time and
lower than the specified upper limit of time value.
Outside a range of time value(--][--)-- Trigger when
the positive or negative slope time of the input signal
is greater than the specified upper limit of time and
lower than the specified lower limit of time value.
Holdoff, coupling, and noise reject can be set in slope
trigger, see the sections " Holdoff", " Trigger Coupling"
and " Noise Reject" for details.
Positive Pulse Width
Trigger Level
Negative Pulse Width
15.5.4 Pulse Trigger
Trigger on a positive or negative pulse with a specified width. Trigger source, polarity
(positive, negative), limit range and time value can be set in the trigger dialog box
Less than a time value(≤)-- Trigger when the positive or negative pulse time of the
input signal is lower than the specified time value. Below is an example of trigger
condition is set to positive pulse width < 100ns
1.800.561.8187
information@itm.com
www.
.com
T3DSO2000A Manual. Page 57