FRLR:The time between the first rising edge of source A and the last rising edge of
source B at the 50% crossing
FRLF:The time between the first rising edge of source A and the last falling edge of
source B at the 50% crossing
FFLR:The time between the first falling edge of source A and the last rising edge of
source B at the 50% crossing
FFLF:The time between the first falling edge of source A and the last falling edge of
source B at the 50% crossing
Skew: Time of source A edge minus time of nearest source B edge

18.4 Trend

After adding a measurement parameter, Trend can be used to observe the change of
the selected measurement value over time.
A. Measurement parameter display area
B. Trend plot display area
C. Extend the range of time. Touch it to expand the time range.
D. Current value
E. Maximum value
F. Minimum value
Press Clear Sweeps button or touch Reset Statistics in the measure dialog box, or touch
the symbol in the statistics display area to clear and restart statistics.
1.800.561.8187
G. Average value
H. Reset statistics