Remote Start/Stop - EBARA ESA70W Instruction Manual

Multi-stage dry vacuum pump
Table of Contents

Advertisement

7.2.2 REMOTE Start/Stop

a) START
Enter the external "MP" start signal input through the control connector. The
MP starts.
In the automatic BP operating mode, the BP can be started/stopped
automatically.
When the BP is operated under external start signal input, input the BP start
signal to the control connector. Please input the start signal when inlet
pressure is below 4000 Pa or less. The BP doesn't start if it doesn't meet this
pressure condition.
The current is indicated on the display during pump operation. For other
status display indications, refer to Table 6.1.
[NOTE]
The pump will not start when an ALARM/WARNING has been generated.
When a START signal is entered, "STARTFAIL" will appear on the display.
b) STOP
When ALTERNATE has been selected, interrupt the external MP start signal
and the pump will stop.
When MOMENTARY has been selected, enter the external MP stop signal
and the pump will stop.
(47)