Troubleshooting interference with other devices or instruments

Troubleshooting interference with other devices or
instruments
When you start an 89600 analyzer, the software attempts to identify other
instruments and devices connected to the computer. This can cause the following
problems:
• Instruments and devices get reset when you start the 89600 analyzer even if the
instrument/device is not controlled by 89600.
• Errors occur on instruments or devices not controlled by 89600 when you start
the 89600 analyzer.
To eliminate these problems, you can instruct the 89600 software to exclude
specific devices from the identification process. To do this, edit the file:
C:/Program Files/Agilent/89600 VSA/VsaVisaTemplate.txt
List the addresses of the devices you want to exclude, then save the file as:
C:/Program Files/Agilent/89600 VSA/VsaVisaConfig.txt
Chapter 16
Diagnostics
233