Performing the Operator's heck
he operator's check should be performed when you receive your analyzer, and any time you
wish to be con dent that the analyzer is working properly. he operator's check does not
verify that the analyzer is performing to speci cations. Its purpose is to provide you with a
high degree of con dence that the analyzer is performing properly if it passes.
he operator's check consists of the following: making a transmission measurement with the
cable that was supplied with your analyzer, making a re ection measurement with the same
cable, and making a measurement with a 50 (or 75 ) termination. Use a known-good load,
such as one from calibration kit HP 85032B/E (50 ) or HP 85036B/E (75 ).
Caution
he preset power level is user-de nable. When pressing
power level to 0 dBm if necessary.
Caution
If you have an P 8730A, please read the following information carefully.
he operator's check, non-automated adjustment procedures, and
troubleshooting procedures all require the HP 8730A tuner analyzer to have the
same preset state as a standard network analyzer. Instead of using the
key, press the following keys to change the preset state of the HP 8730A tuner
analyzer to the preset state of a standard network analyzer:
SYSTEM OPTIONS
ervice Utilities Next Menu Next Menu
Inst tate Utilities Preset to 8711C
1-6
Performance T ests
ervice
5
, change the
PRESET
5
.
PRESET
5