A4 - Acoustic Alarm Silencing - Siare Falco 202 User Manual

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

A4 - Acoustic alarm silencing

5.1.5
The soft key, during the normal operating phase of ventilator,
allows the silencing of active acoustic alarm.
Pushing the RESET button will stop the acoustic alarm
for a defined time.
During the alarm silencing the alarm text is showed on
the panel.
Pushing further the RESET button will cancel the alarm
text only if the alarm conditions is disappeared.
If during the alarm silencing a new high priority alarm
occur, the alarm silencing is cancelled and the acoustic
signal and the visual texts are activated again.
Falco 202
It's possible to change the alarm settings even when the alarms are
activated.
After changing an alarm setting, the relevant sign is lighted and the status
icon will flash for a defined time.
WARNING !!
Risk of injury for the user / patient
The user should never stop checking the patient conditions during the
alarm silencing.
5-7