5.1.5 A4 - Acoustic alarm silencing

During the normal operating phase of lung ventilator, it is
possible to silence the active acoustic alarm.
By pressing the Alarm Reset soft key.
By selecting the bell evidencing the active state of the
alarm.
WARNING !!
The User should never stop checking the patient conditions during the alarm
silencing.
NOTE
Falco 202 Evo
It's possible to change the alarm settings even when the alarms are
activated.
After changing an alarm setting, the relevant sign is lighted and the
status icon will flash for a defined time.
Risk of injury for the user / patient
The acoustic alarm silencing is active and will stop the acoustic alarm for
a defined time.
During the alarm silencing the alarm text is showed on the panel.
A new alarm silencing will cancel the alarm text only if the alarm
conditions is disappeared.
If during the alarm silencing a new high priority alarm occur, the alarm
silencing is cancelled and the acoustic signal and the visual texts are
activated again.
5-7