5.1.4 A3 -
General information area
This area shows the battery charge level and the power supply status (present/absent).
Green "BATTERY" symbol, indication of the battery charge level:
with fix symbol the battery is complete charge;
with flashing symbol the battery is in charging phase.
Green "PLUG" symbol, indication of the main power supply presence.
Red "PLUG" symbol, flashing: indication of power failure.
The battery charge level is evidenced by the presence of colored "notches"
within the symbol, where each notch represents the 25% charging level.
Green flashing symbol, 2 notches: it indicates that the charge level
of the battery is at 50% - the relevant alarm is active.
Orange flashing symbol, 2 notches: it indicates that the battery
charge level is at 50% - the relevant alarm is active.
Green flashing symbol, 1 notch: it indicates that the battery charge
level is at 25% - the relevant alarm is active.
Red flashing symbol, 1 notch: it indicates that the battery charge
level is at 25% - the relevant alarm is active.
The colour of the last flashing "notch" is red (high priority alarm): this
extremely serious alarm condition indicates that the battery is almost
completely low.
Red color signal, flashing: indication of main power failure.
The green "PLUG" symbol: indication of the presence of main
power supply.
The power failure alarm is reported both by the visual signals of the
corresponding message in the alarm area and the high priority red bell as
well as by the flashing electrical power symbol.

5.1.5 A4 - Acoustic alarm silencing