5.1.3 A2 - MAIN MENU / ALARMS parameters

This monitor area, through the encoder knob ( MAIN MENU - ALARMS ), allows to display
and set the Min and Max. alarms values.
Low Pressure
High Pressure
Low Resp. Rate
High Resp. Rate
Low Insp. Vti
High Insp. Vti
Low FiO2
High FiO2
Max. Leak (l/min)
Power Failure
Falco 101
Risk of injury for the user / patient
The ventilators used in the same health environments can have different
preset configurations of alarm limits.
Verify that the preset alarm limits are appropriate for the new patient and
adjust the alarm limits on values suitable to the new condition of use.
From 2 to 20cmH2O
From 2 to 65 cmH2O
From 1 to 10 bpm
From 1 to 70 bpm
From 0 to 249 ml (10 ml steps)
From 5 to 2500 ml (10 ml steps)
From 20 to 98 %
From 21 to 100 %
From 1 to 150
Enable / Disable
5-5