A2 - Main Menu / Alarms Parameters - Siare Falco 202 User Manual

Lung ventilator turbine-driven ventilation
Table of Contents

Advertisement

A2 - MAIN MENU / ALARMS parameters

5.1.3
This monitor area, through the encoder knob ( MAIN MENU - ALARMS ), allows to display
and set the Min and Max. alarms values.
Low Pressure
High Pressure
Low Resp. Rate
High Resp. Rate
Low Exp. Vte
High Exp. Vte
Low Exp. VM
High Exp. VM
Low PEEP
High PEEP
Low FiO2
High FiO2
Gas Sensor : Low EtCO2
Gas Sensor : High EtCO2
Power Failure
Falco 202
WARNING !!
Risk of injury for the user / patient
The ventilators used in the same health environments can have different
preset configurations of alarm limits.
Verify that the preset alarm limits are appropriate for the new patient and
adjust the alarm limits on values suitable to the new condition of use.
From 2 to 60 cmH2O
From 20 to 81 cmH2O
From 1 to 11 bpm
From 13 to 150 bpm
From 0 to 1400 ml (10 ml steps)
From 1 to 3000 ml (10 ml steps)
From 0 to 20 L
From 1 to 100 L
From 0 to 19 cmH2O
From 1 to 51 cmH2O
From 20 to 98 %
From 21 to 100 %
From 0.4 to 12.9 %
From 0.5 to 13.0 %
Enable / Disable
5-5