1.
Remove the reservoir bottle caps.
2.
Screw the reservoir bottles onto the instrument, as shown below.
3.
For each reservoir bottle, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the bottle.
To install low-volume vials:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
If a standard reservoir bottle is fitted, remove it.
2.
Screw the low-volume adaptor into the manifold and tighten it.
3.
Screw the low-volume vial into the adaptor.
4.
For each low-volume vial, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the vial.
The reservoir bottles can be contaminated with
Preparing the IntelliStart Fluidics system
TP03410
2-7