5.
Clear an area, lay out the soft cloth or mat, and lay the source enclosure
on its face.
6.
Use the Phillips (cross-head) screwdriver to remove the 2 source
enclosure, release-handle screws, and remove the handle.
Source enclosure
cover screws
Lamp-drive cover
screws
Release handle
7.
Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the remaining 2 lamp-drive cover screws, which were hidden by
the release handle.
8.
Slide the cover off the lamp drive.
9.
Use the 3-mm Allen wrench to remove the 4 source enclosure cover
screws.
10. Ease the source enclosure cover over the lamp drive assembly.
4-110 Maintenance Procedures
Take care to lay the source enclosure on a smooth sur-
Caution:
face. Laying it face-first on a hard object or other protrusion can
smash the glass window.
TP03204