ESI mode
The following sections explain how to install and remove an ESI probe. For
further details on running ESI applications, see
Installing the ESI probe
Required material
Chemical-resistant, powder-free gloves
To install the ESI probe:
Warning:
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
prepared for working on the source before commencing this procedure.
1.
Prepare the instrument for working on the source (see
Warning:
handle it with care.
2.
Remove the protective sleeve, if fitted, from the ESI probe tip.
3-2
Changing the Mode of Operation
The LC system connections, ESI probe, and source can
To avoid electric shock, ensure that the instrument is
The ESI probe tip is sharp. To avoid puncture wounds,
page
1-9.
page
4-7).