3.
Use the 3-mm Allen wrench to unscrew and remove the 4 screws
securing the PEEK ion block support to the adaptor housing, and
remove the ion block support.
Securing screws
4.
If any of the O-rings show signs of deterioration or damage, dispose of
them in accordance with local environmental regulations.
4-56
Maintenance Procedures
PEEK ion block
support
The O-rings can be contaminated with
Warning:
biohazardous and/or toxic materials. Dispose of them
according to local environmental regulations.
Adaptor housing
Ion guide
assembly