Waters Xevo TQD Operator's, Overview And Maintenance Manual page 72

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

To fit a borosilicate glass capillary (nanovial):
Caution:
handling them. Always hold their blunt end, never the sharp end, which
can easily be damaged.
Warning:
fully retracted from the source before beginning this procedure.
1.
Loosen the stage retaining screw.
2.
Pull the stop screw to release the stage.
3.
Slide the stage out of the NanoFlow source enclosure and remove the
magnetic cover.
4.
Unscrew the retaining screw, and lift the sprayer from the stage.
5.
Unscrew the union from the end of the sprayer assembly.
6.
Remove the existing capillary from the sprayer.
3-20
Changing the Mode of Operation
To avoid lacerations, puncture injuries, and
Warning:
possible contamination with biohazardous and toxic samples,
do not touch the sharp end of the capillary.
Capillaries are extremely fragile. Take great care when
To avoid electric shock, ensure that the NanoFlow stage is
Capillary
Union

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents