Safety goggles
To install the new capillary
Warning:
the probe with care.
1.
Use the sharp knife or PEEK tubing cutter to cut a piece of red PEEK
tubing approximately 60 cm (24 inches) long.
Requirement:
horizontal axis).
2.
Insert one end of the red PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Rationale:
capillary.
3.
Fit the UNF coupling to the new capillary.
4-74
Maintenance Procedures
:
The probe and source components can be contaminated
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
The ESI probe tip is sharp. To avoid puncture wounds, handle
Cut the tubing squarely (that is, perpendicular to its
Doing so ensures a minimum dead volume when fitting the
Probe inlet connector
PEEK tubing
TP02671