Appi/Apci Source-Replacing The Appi Lamp Drive Seals; Removing The Appi Lamp Drive Assembly Seals - Waters Xevo TQD Operator's, Overview And Maintenance Manual

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

2.
Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure
3.
Use methanol or isopropyl alcohol, applied to the lint-free cloth, to
carefully clean the lamp window.
4.
Close the source enclosure.
5.
Slide closed the instrument's source interface door.
APPI/APCI source—replacing the APPI lamp drive seals
The following APPI lamp drive assembly O-rings must be renewed at
intervals of no greater than 1 year:
UV lamp bulb sealing O-ring
Mounting shaft O-rings
UV lamp mounting flange O-ring
An automatic pressure test runs each time you close the source enclosure
Tip:
and when the instrument starts.

Removing the APPI lamp drive assembly seals

Required materials
Chemical-resistant, powder-free gloves.
Combined 2.5-mm Allen wrench and cone extraction tool.
4-108 Maintenance Procedures
The probe, source, and lamp bulb can be hot. To avoid
Warning:
burn injuries, take great care while working with these compo-
nents.
To avoid eye injury from UV radiation, ensure that the
Warning:
APPI lamp is extinguished before carrying out this procedure.
To avoid leaking of biohazardous or toxic materials,
Warning:
ensure the integrity of the source exhaust system. The APPI lamp
drive assembly O-rings listed below must be renewed at intervals
not exceeding 1 year, exactly as described in this section.