Warning:
take great care while working with these components.
2.
Remove the probe from the currently installed source.
If you are removing an ESI probe, see
If you are removing an IonSABRE II probe, see
3.
Remove the existing source enclosure (see
4.
Install the combined APPI/APCI source enclosure (see
5.
Install the corona pin (see
6.
Connect the APPI drive cable to the instrument's front panel connector.
7.
Connect the HT cable to the instrument's front panel connector.
Caution:
ensure that the lamp assembly does not touch the corona pin when
the source enclosure door is closed.
8.
Install the IonSABRE II probe to the source, and ensure that it is
working correctly (see
An automatic pressure test runs each time you close the source
Tip:
enclosure and when the instrument starts.
3-14
Changing the Mode of Operation
The probe and source can be hot. To avoid burn injuries,
page
4-12).
To prevent damage to the corona pin and lamp assembly,
page
3-6).
page
3-4.
page
3-8.
page
4-8).
page
4-11).