Waters Xevo TQD Operator's, Overview And Maintenance Manual page 159

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

15. Carefully push the PEEK union/UNF coupling assembly and capillary
into the probe assembly so that the locating pin on the UNF coupling is
fully engaged in the locating slot at the head of the probe assembly.
UNF coupling locating pin
16. Fit the nebulizer adjuster knob to the PEEK union/UNF coupling
assembly.
17. Finger-tighten the nebulizer adjuster knob onto the probe assembly.
18. Fit the probe gasket and end-cover to the probe assembly.
19. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 3 screws retaining the probe end-cover.
20. Return the combined 2.5-mm Allen wrench and cone extraction tool in
its storage location on the source adaptor housing.
Caution:
• When handling the probe heater, take great care to grip the
heater so as not to damage its electrical wiring.
• Take great care not to damage the probe heater's electrical
connections, capillary sleeve, or capillary when fitting the
heater over the capillary sleeve.
21. Fit the probe heater (see
22. Fit the probe to the instrument (see
23. In the Instrument Console, click API
desolvation gas flows.
Probe assembly locating slot
page
4-86,
step 1
page
Replacing the IonSABRE II probe sample capillary
through
step
3).
3-6).
to start the probe and
4-83