Waters Xevo TQD Operator's, Overview And Maintenance Manual page 157

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

Needle-nose pliers
7-mm wrench
Combined 2.5-mm Allen wrench and cone extraction tool
Red PEEK tubing
LC pump
HPLC-grade (or better) 1:1 acetonitrile/water
Capillary
Sharp knife or PEEK tubing cutter
Safety goggles
To install the new capillary:
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Use the sharp knife or PEEK tubing cutter to cut a piece of red PEEK
tubing approximately 60 cm (24 inches) long.
Requirement:
horizontal axis).
2.
Insert one end of the red PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Doing so ensures a minimum dead volume when fitting the
Rationale:
capillary.
The probe and source components can be contaminated
Cut the tubing squarely (that is, perpendicular to its
Probe inlet connector
PEEK tubing
Replacing the IonSABRE II probe sample capillary
TP02671
4-81