Waters Xevo TQD Operator's, Overview And Maintenance Manual page 155

Hide thumbs Also See for Xevo TQD:
Table of Contents

Advertisement

3.
Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 3 screws retaining the probe end-cover.
4.
Remove the end cover and gasket.
Nebulizer adjuster knob
5.
Unscrew and remove the nebulizer adjuster knob.
6.
Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
End-cover retaining screws
Replacing the IonSABRE II probe sample capillary
Gasket
End-cover
4-79