To remove the source enclosure:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Prepare the instrument for working on the source (see
Warning:
take great care while working with these components.
2.
Remove the probe from the source.
If you are removing an ESI probe, see
If you are removing an IonSABRE II probe, see
3.
Disconnect the probe adjuster and options cables from the instrument's
connectors.
Warning:
take great care while working with the source enclosure open if a
corona pin is fitted.
4.
Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
The source components can be contaminated with
The probe and source can be hot. To avoid burn injuries,
The corona pin tip is sharp. To avoid puncture wounds,
Removing and refitting the source enclosure
page 4-7).
page
3-4.
page
3-8.
4-9