To clean the sample cone and cone gas nozzle
Warning:
contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing
this procedure.
Warning:
extreme care, use a fume hood and suitable protective
equipment.
The sample cone is fragile. Never place it on its tip; always
Caution:
place it on its flanged base.
1.
If the sample cone contains debris, place a drop of formic acid on its
aperture.
2.
Immerse the sample cone, cone gas nozzle, and cone gas nozzle handle in
separate glass vessels containing 1:1 methanol/water.
If the components are obviously contaminated, use 45:45:10
Tip:
methanol/water/formic acid.
3.
Place the vessels in the ultrasonic bath for 30 minutes.
4.
If you used formic acid in the cleaning solution, do as follows:
a.
Rinse the components by immersing them in separate glass vessels
containing water and then place the vessels in the ultrasonic bath
for 20 minutes.
b.
Remove any residual water from the extraction cone by immersing
it in a glass vessel containing methanol and then place the vessel in
the ultrasonic bath for 10 minutes.
Caution:
chemical-resistant, powder-free gloves for the rest of this
procedure.
5.
Carefully remove the components from the vessels, and blow-dry them
with inert, oil-free gas.
The sample cone and cone gas nozzle can be
Formic acid is extremely corrosive and toxic. Work with
To avoid recontaminating the components, wear clean,
Cleaning the sampling cone assembly
4-33