B.7

Connecting the liquid waste line

B.7.1

Required material

Chemical-resistant, powder-free gloves
To connect the liquid waste line:
Warning:
To prevent personal contamination with biohazards or toxic materials, always wear
chemical-resistant, powder-free gloves while performing this procedure. The waste line and
connection can be contaminated.
1.
Place a suitable waste container below the instrument.
Notice:
drain cup when attaching or removing the waste line.
2.
Slide a drain line over the barbed fitting of the drain (located at the bottom of the
instrument).
Warning:
drain line does not crimp or bend, impeding flow to the waste container.
waste container is emptied before the lower end of the drain tube is covered by waste
solvent.
3.
Route the waste line to the waste container. If necessary, shorten the waste tube so that its
end is above the surface of the waste solvent.
To avoid distorting the drip tray or causing the drain cup to leak, restrain the
Drain line
Barbed fitting
To prevent leakage of biohazards or toxic materials, ensure that the
September 7, 2016, 715003560IVD Rev. C
Drain cup
TP02684