To reboot the instrument's software:
1.
Ensure that MassLynx software is closed.
2.
Insert a short length (about 7.5 cm) of PEEK™ tubing, or a similar object, into the reset
button aperture at the top, right-hand side of the instrument's front panel.
3.
Remove the PEEK tubing from the reset button aperture.
4.
Wait until the rebooting sequence ends before starting the MassLynx software.
Tip:
An alert sounds at the end of the rebooting sequence.
2.4

Leaving the mass spectrometer ready for operation

Leave the mass spectrometer in Operate mode except in the following cases:
When performing routine maintenance
When changing the source
When leaving the mass spectrometer unused for a long period
In these instances, put the mass spectrometer in Standby mode. See the online Help for details.
2.4.1

Emergency shutdown of the mass spectrometer

Warning:
below. The instrument's power switch does not isolate it from the main power supply.
Notice:
To avoid losing data, reboot the instrument as described on
lost when you perform an emergency shutdown.
To shut down the instrument in an emergency:
1.
Operate the power button on the front of the instrument.
2.
Disconnect the power cable from the instrument's rear panel.
Reset button aperture
To avoid electric shock, isolate the instrument observing the procedure outlined
September 7, 2016, 715003560IVD Rev. C