Constant Neutral-Loss Mode - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

scans (in MS mode) the background for all other components. RADAR mode enables fast
characterization of potential matrix effects, providing a platform for more robust method
development.
1.6.5.1
Typical application
You typically use RADAR mode during method development before performing MRM or PICS
analyses to quantify known analytes in complex samples:
Drug metabolite and pharmacokinetic studies
Environmental analysis: for example, for pesticides and herbicides
Forensic or toxicologic analysis: for example, screening for target drugs in sports
MRM analysis with no associated RADAR or PICS operation does not produce a spectrum
because only one transition is monitored at a time. As with SIR mode, however, a chromatogram is
produced.
1.6.6

Constant neutral-loss mode

Constant neutral-loss mode detects the loss of a specific neutral fragment or functional group from
an unspecified precursor or precursors.
The scans of MS1 and MS2 are synchronized. When MS1 transmits a specific precursor ion, MS2
"looks" to see whether that precursor loses a fragment of a certain mass. If it does, the loss
registers at the detector.
In constant neutral-loss mode, the spectrum shows the masses of all precursors that actually lost a
fragment of a certain mass.
Figure 1–7:
1.6.6.1
Typical application
You typically use constant neutral-loss mode to screen mixtures for a specific class of compound
that is characterized by a common fragmentation pathway, indicating the presence of compounds
containing a common functional group.
Constant neutral-loss mode
MS1
Scanning
(synchronized with MS2)
September 7, 2016, 715003560IVD Rev. C
Page 28
Collision cell
MS2
Fragmenting precursor
Scanning
ions and passing all
(synchronized with MS1)
masses

Advertisement

Table of Contents
loading

Table of Contents