Operation
Flow Rate
Calibration

Flow Rate Calibration

3-26
MIE ADR-1500 Instruction Manual
From the RH Offset screen, press the
offset.
Press the
ENTER
Offset screen.
RH OFFSET
* 5.0
The Flow Rate Calibration screen allows the user the option to perform a
multi-point inlet flow calibration.
From the Calibrate Menu, press the
key to the Flow Rate Cal screen.
FLOW RATE CAL
The Flow Rate Calibration Edit screen allows the user to calibrate the six
flow control settings and define the flow rate at each flow control setting.
Edit
This will build a table of the six calibration point to be used for
interpolating the flow control setting to achieve the flow rate set in the
Configure menu's "Flow Rate Edit".
The flow rate calibration is accomplished in a sequence of six displays for
six flow calibration points, setting the six flow control settings for the
corresponding six defined flow rate setting. The
used to navigate through the edit fields, while the ▲ and ▼ keys are used
to increase and decrease the values. The asterisk (*) appears to the left of the
field being edited.
For factory settings, the flow control adjust is adjusted for flow calibration
points (1, 2, 3, 4, 5, and 6) to produce corresponding defined flow rates
(1.00, 1.50, 2.00, 2.50, 3.00, and 3.50 liters per minute) and the defined
flow rates are kept at the factory default setting. The user may, however,
redefine the defined flow rate for each flow calibration point to get a more
accurate calibration. However, the defined flow rate values must increase in
value as the flow calibration point increases.
In the following screen there are two values shown; ADJ = 188 and LPM =
1.00. In this screen the 188 is the pump speed control value used to achieve
1.00 L/min. The user may adjust the motor speed to achieve 1.00 L/min or
ENTER
key to save the user RH offset and return to the RH
42.3 %
ENTER
key and edit the RH
key and scroll the ▲ or ▼
and
keys are
ENTER
ESC