Installing The Manual Injection Port - Teledyne ACCQPrep HP125 Installation And Operation Manual

Table of Contents

Advertisement

Figure 2-13 Turn the valve to the Prime position
2.3 Installing the Manual
Injection Port
Figure 2-14 ACCQPrep with Manual Injection Luer Port installed
• If the solvent select option is not installed, the system
will immediately prime both solvent lines.
• If the solvent select option is installed, a dialog box
opens when the valve is switched to the P
The dialog box allows you to select which of the A
solvents and B solvents will be primed.
• If you have not already assigned names to the A and B
solvents, they can be named at this time using the drop
downs. If the name you want isn't listed, it can be added
in the C
ONFIGURATION
completed at a later time.
3. Once priming is complete, a new screen will appear direct-
ing you to turn the valve from the P
R
position.
UN
Note
The system will prime both pumps simultaneously. There
is no method to prime A and B individually.
Manual sample injection can be performed with the system even
when equipped with an AutoInjector or AutoSampler.
If no AutoInjector or AutoSampler is installed, a luer injection
port (209900002) is included in the accessory package. To install,
place the fitting into the injection valve and tighten the nut
snugly (Figure 2-14).
ACCQPrep HP125
Section 2 Installation
RIME
screen. This action can be
position to the
RIME
position.
2-9