A plug has occurred in the PurIon fluid lines. The separa-
tion can be continued without PurIon detection or contin-
ued if the plug is corrected. Error 310
The PurIon inlet cone may be plugged which could pre-
vent detection. Continued operation will not cause dam-
age. Please contact a qualified service technician to
clean the cone; Pirani Pressure: [s1] mbar. Error 315
A plug has occurred in the PurIon fluid lines. The plug
must be corrected to allow continued operation. Error
316.
The ionization probe (ESI or APCI) isn't fully seated into
the ion source housing. Please ensure the probe is fully
seated, then press OK to continue. Error 317.
The fluid interface pressure is too low. Error 325
Table 6-1 Common PurIon
Error Codes and Resolutions
This error message is displayed if a plug is detected dur-
ing a separation.
During normal operation, the Pirani pressure should be
>1.5E-3 mbar. Anything less is an indication of either par-
tial or complete plugging of the capillary cone entrance to
the vacuum area. See cleaning the cone capillary (6.8.3).
This error occurs if the tubing is plugged while the
method development screen is in use. See cleaning the
probe capillary (6.8.3).
On PurIon systems (not PurIon S or PurIon L), the probe
isn't seated properly. Loosen the thumbscrew on the front
of the ion source housing, push down on the probe, then
tighten the thumbscrew.
This could be due to lack of carrier fluid, loss of pump
prime, or leakage.
ACCQPrep HP125
Section 6 Maintenance
6-27