Teledyne ACCQPrep HP125 Installation And Operation Manual page 39

Table of Contents

Advertisement

AutoSampler control not visible on the ACCQPrep
software – The ACCQPrep only loads the AutoSampler control
features if it can detect the presence of the AutoSampler Check
the AutoSampler pilot light (upper right corner of AutoSampler)
to determine the next step.
• If the light is off:
· Verify that the AutoSampler power switch is on. This
is located on the rear panel of the AutoSampler.
· Check that the AutoSampler power cord connects
from the ACCQPrep to the power connection on the
rear panel of the AutoSampler.
• If the light is blue the AutoSampler is functioning but
not communicating with the ACCQPrep:
· Check the USB cable connection from ACCQPrep to
the square USB connector along the rear, bottom
edge of the AutoSampler (Figure 2-22).
· Check the USB cable connection from the lower, rear
panel of the AutoSampler to the USB connector on
the upper left corner of the rear panel (while viewing
the AutoSampler from the rear panel.
· Recycle the power on the ACCQPrep. The PeakTrak
software only checks for the presence of the AutoS-
ampler on power up. If the AutoSampler is powered
up after the ACCQPrep, the software will not detect
the presence of the AutoSampler.
• If the light is red the AutoSampler is not able to reach
the home position:
· The sampler arm is not able to move to the left rear
corner to home. Verify that there is nothing blocking
its motion.
· Inspect the flag on the rear of the probe mechanism
to ensure the flag is perpendicular to the rear panel.
It must enter into a receptacle to reach the home
position. Its mounting screw may be tightened, but
too much torque could break the plastic screw.
· The sample probe is unable to reach the top of its
travel. Verify that there is nothing blocking its
motion.
ACCQPrep HP125
Section 2 Installation
2-23