Waste Port Mounting Plate; Deuterium Lamp Location - Teledyne 10 Installation And Operation Manual

Teledyne isco, inc. installation and operation guide switch
Table of Contents

Advertisement

BioOptix™ 10
Section 4 System Maintenance
Figure 4-5 Waste port mounting plate
Lamp Mounting Screws
Twist Lock Clamp
Figure 4-6 Deuterium lamp location
4-6
Hot surfaces may cause burns. The UV lamp and nearby sur-
faces are hot.
To remove the lamp:
1. Remove the two screws that secure the waste port mount-
ing plate (Figure 4-5). The optical assembly is located
behind this plate.
2. Disconnect the wiring connector.
3. Release the twist lock cable clamp.
4. Remove the three lamp mounting screws.
5. Carefully pull the lamp straight out from the optical
assembly.
CAUTION