Pump Installation - Teledyne ACCQPrep HP125 Installation And Operation Manual

Table of Contents

Advertisement

ACCQPrep HP125
Section 2 Installation

2.2.1 Pump Installation

Figure 2-2 Location of notches on the pump sled
2-2
2. Push the arm to the left or right to remove the spacer.
3. Store the bracket and spacer. These parts should be rein-
stalled if the system must be shipped again.
When you are ready to place the system, be sure it is situated in
such a way that the power cord is accessible to disconnect in case
of malfunction.
Note
Be sure to move the two electrical connections and solvent
tubing out of the way before sliding the pump assembly
into the pump compartment.
To install the pump:
1. Remove the two pump retaining screws in the pump com-
partment.
2. Align the pump assembly horizontally with the two notches
facing the opening of the pump compartment (Figure 2-2).
3. Slide the pump assembly into the pump compartment. It
may require a little force since it is metal on metal sliding
together. If the force seems excessive, ensure there is noth-
ing obstructing the pump assembly. Be sure to keep the
electrical connections and tubing out of the way so they do
not get pushed inside of the compartment (Figure 2-3).
Note
Be sure to observe the top of the pump assembly as it is being
inserted into the system. Make sure no wires from the case or
pump assembly are being snagging.