Teledyne ACCQPrep HP125 Installation And Operation Manual page 30

Table of Contents

Advertisement

ACCQPrep HP125
Section 2 Installation
Auto Sampler Probe
Wash Pump
1 – Wash Inlet
2 – From Wash Station
3 – To Wash Station
4 – Waste
Not Used
Waste line
Wash station
supply
Needle Wash
Figure 2-22 Tubing diagram of the AutoSampler connected
to the ACCQPrep (See Appendix A, Figure A-1
for a larger diagram)
2-14
Auto Injector
Optional
Column Valve
To bottom
Optional
column port
To Injection
Valve
Not Normally Used
10. Place the AutoSampler Module on the right side of the
ACCQPrep.
11. Connect the AutoSampler tubing labeled "To ACCQPrep"
to the connection on the ACCQPrep's rear panel labeled "To
AutoSampler". The length of this tubing is critical to con-
trol the delay volume for fraction collection on the AutoS-
ampler. The overall length of the tubing to the diverter
valve should be 67" (170 cm).
12. Route the "To Waste" tubing to a suitable waste container
or tee to either the Wash Waste Line from the AutoSam-
pler Module (from step 8 above or seen in Figure 2-22)
and/or the ACCQPrep waste line and injection valve waste
line with the supplied tee(s).
Auto
Auto
External
Sampler
Sampler
Detector
Power
Power
Ethernet
USB
USB
Check Valve
Sample
Main
Main
A1
A1
A2
A2
A3
A3
Power
Power
Optional
External Gas
Not Normally Used
Solvent Level Sensing
Solvent Level Sensing
B1
B1
B2
B2
B3
B3
Drain
Drain
Circuit Breaker
Needle Wash
Inlet Line
Waste
To
To
Auto
Auto
Sampler
Sampler
Wash Station
Fraction Collection
Solvent Level
Sensing
Solvent Lines
Do not
extend length.
See Step 11
Waste
Waste
Level
Level
Waste
Waste
Sense
Sense
for details.
Solvent Inlet
Solvent Inlet
J
A1
A1
A2
A2
Solvent inlet
A3
valves
optional
If not used,
A
A
B
B
connect directly
IN
IN
IN
IN
to A and B
B3
B3
B1
B1
B2
B2
B4
B4
WASTE
Repeat tubing for A1 Solvent as needed
A1 Solvent