ACCQPrep HP125
Section 1 Introduction

1.5 Specification

Dimensions of ACCQPrep
(Footprint without column or column cover
installed)
Dimensions of ACCQPrep AS 2x1
Weight
(Including AutoInjector, Column Selector
Valve Module, Solvent Selector Valve Mod-
ule, and column cover)
Weight of ACCQPrep AS 2x1
User Interface
Power
Line Frequency
Ambient Temperature
Humidity (when connected to power)
Flow Rate Range
Flow Rate Accuracy (tested with water at 6.9
bar or 100 psi)
Pressure Limit
Pressure Accuracy
Gradient Formation
Gradient Accuracy:
Peak Detection Modes
Flow Cell Pathlength
UV Detection Wavelength
Wavelength Accuracy
Fraction Accuracy
Optional ELSD
Gas Inlet Pressure
Gas Consumption
Spray Chamber Temperature
Drift Tube Temperature
1-4
Table 1-1 ACCQPrep HP125
27.5 x 14.0 x 20.0 in (69.9 x 35.6 x 50.8 cm)
18.9 x 14.0 x 19.3 in (48.0 x 35.6 x 49.0 cm)
93.2 lbs (42.3 kg)
106 lbs (48 kg) with optional Evaporative Light Scattering Detector
(ELSD)
27.3 lbs (12.4 kg)
15" touchscreen
Input voltage range from 100 to 240 VAC,
300 VA maximum.
Line cord is the disconnect device. Power connection is via IEC
60320 C14 power inlet.
50/60 Hz
20 to 40 °C (maximum temperature must be at least 15 °C below the
boiling point of the solvent)
90% relative humidity maximum at 20 to 40 °C (non-condensing)
1-125 mL/min, ELSD peak collection limited to 4-125 mL/min
± 2%
414 bar (6000 psi) derated linearly from 6000 psi at 75 mL/min to
4000 psi at 125 mL/min
5% of full scale
Binary gradient 2 solvent inlets. Optional Solvent Selector Valve Mod-
ule expands to 3 solvents each for A & B
±1% of full scale
Slope or threshold
0.3 mm, ±25% (Standard, other pathlengths available)
200 to 400 nm, optional 200 to 800 nm UV-Vis
±5 nm
±[2mL + (flow rate ÷ 60)]
60 to 70 psig
<2.5 SLPM
Setting range: 10 to 60 °C
limited to minimum of 5 °C below ambient
Setting range: 30 to 90 °C
Must be 5 °C above spray chamber temperature Minimum tempera-
ture is 5 °C above ambient Maximum temperature is 60 °C above
ambient