Fluke Calibration 5322A Operator's Manual page 158

Multifunction electrical tester calibrator
Hide thumbs Also See for 5322A:
Table of Contents

Advertisement

5322A
Operators Manual
Wait-to-Continue command
*WAI
Prevents the Product from executing any further commands or queries until all
previous remote commands have been executed.
Reset
*RST
This command resets the Product to its initial status.
Test operation
*TST?
This command launches an internal self-test. Return the self-test result (0 for
pass or 1 for fail).
Read Status Byte
*STB?
This query returns number in range 0 to 255 with information about content of
register STB, which carries the MSS bit status.
Service Request Enable
*SRE <value>
This command sets condition of the Service Request Enable register. Since bit 6
is not used, the maximum value is 191.
Read Service Request Enable
*SRE?
Returns the Service Request Enable Register number.
Read Event Status Register
*ESR?
Returns the contents of the Event Status Register and clears the register.
Set Event Status Enable
*ESE <value>
This command programs the Event Status Enable register bits. Parameter value
is number in range 0 – 255.
Read Event Status Enable
*ESE?
Returns the Event Status Enable register.
Clear status
*CLS
This command clears the Event Status Register and the Status Byte Register
except the MAV bit and output queue. Output line is not reset.