[SOUR]:SAF:GBR:LOWC(?) <CPD> { ON | OFF }
Description: This command selects the Low Current feature of the Ground
Parameters: <CPD>
Example:
Query:
[SOUR]:SAF:GBR:SRES(?) <DNPD>
Description: This command sets the series resistance of connection cables. If
Parameters: <DNPD>
Example:
Query:
[SOUR]:SAF:GBR:CURR?
Description: This command returns the measured current flowing through the
Query:
[SOUR]:SAF:GBR[:CURR]:LIM?
Description: This command returns the maximum current allowed flowing
Query:
106
Bond Resistance mode. If not already selected, this command
switches the Product to the Ground Bond Resistance function.
SAF:GBR:LOWC ON
SAF:GBR:LOWC?
not already selected, this command switches the Product to the
Ground Bond Resistance function.
SAF:GBR:SRES 0.072
SAF:GBR:SRES?
selected resistance. The Product returns the current value using
exponential format. Example: 2.5A is returned as 2.500e+00. If not
already selected, this command switches the Product to the Ground
Bond Resistance function.
SAF:GBR:CURR?
through the selected resistance. The Product returns the current
value using an exponential format. Example: 10 A is returned as
10.000e+00. If not already selected, this command switches the
Product to the Ground Bond Resistance function.
SAF:GBR:LIM?
ON activates Low Current feature
OFF deactivates Low Current feature
Low Current feature activation
Product returns ON or OFF.
Parameter represents the value of series
resistance in Ω. The Specifications (online
at www.Flukecal.com) lists the acceptable
ranges.
Product returns the value of series
resistance using exponential format.
Example: 0.072Ω is returned as
0.072e+00.