5322A
Operators Manual
Calibrator Functions
The Product performs output and measurement functions.
Output Functions
•
•
•
•
•
•
Measurement Functions
•
•
•
•
•
Function Descriptions
The subsequent sections describe various functions of the Product. Unless
otherwise noted, the descriptions apply to all models of the 5322A.
Insulation Resistance
When using the 5322A for insulation resistance calibration, the Product acts as a
high-resistance source from 10 kΩ to 10 GΩ with 4½ digit resolution. A single
value 100 GΩ selection is available as well. Depending on the selected
resistance value, maximum applied test voltages range from 50 V to 1500 V
peak.
The 5322A/5 has a 5 kV high-resistance source with a fully-programmable
resistance range from 10 kΩ to 100 GΩ. Maximum test voltage for this option is
from 50 V to 5500 Vpk MAX, depending on setup resistance value.
Earth Resistance and Continuity
The Product sources low resistance values from 100 mΩ to 10 kΩ with 3½ digits
of resolution. This function is used in either 2-wire or 4-wire mode to calibrate
continuity testers and earth resistance testers that source currents from 5 mA to
700 mA.
6
Insulation resistance
Earth resistance and continuity
Loop, line, and ground bond resistance
RCD and Ground Fault Circuit Interrupter (GFCI) tests
Leakage current source
AC/DC voltage generation (5322A/VLC only)
AC/DC voltage and current measure
AC Power, including phase
Hipot voltage distortion, ripple coefficient
Flash test voltage, Class I and Class II
Load current