5322A
Operators Manual
[SOUR]:SAF:VOLT:LOW(?) <CPD> {FLOat | GROund}
Description: This command connects or disconnects the Voltage Calibrator Lo
Parameters: <CPD>
Example:
Query:
[SOUR]:SAF:MET:FUNC(?) <CPD> {DC | AC}
Description: This command selects the dc or ac multimeter mode. This
Parameters: <CPD>
Example:
Query:
[SOUR]:SAF:MET:PROB(?) <CPD> { OFF | 5KV | 10KV | 40KV }
Description: This command selects the HV internal or external high voltage
Parameters: <CPD>
Example:
Query:
[SOUR]:SAF:MET:CURR?
Description: This command returns the measured current. The Product returns
Query:
140
terminal to/from GND terminal. This command switches the Product
into the Voltage Calibrator function mode.
SAF:VOLT:LOW GRO Grounds the Voltage Calibrator Lo output
SAF:VOLT:LOW?
command switches the Product into the Meter function mode.
SAF:MET:FUNC AC
SAF:MET:FUNC?
probe for the voltmeter. This command switches the Product into
the Meter function mode.
SAF:MET:PROB 10KV
SAF:MET:PROB?
the current value using exponential format. Example: 215 mA is
returned as 215.00e-03. This command switches the Product into
the Meter function mode.
SAF:MET:CURR?
GROund grounds the Voltage Calibrator
Output.
FLOat floats the Voltage Calibrator output
from ground.
terminal.
The Product returns GRO when the output
is grounded or FLO when floating.
DC sets to the dc multimeter mode.
AC sets to the ac multimeter mode.
Sets to the ac multimeter mode.
The Product returns ac or dc.
OFF – multimeter with no probe (V terminal)
5KV – multimeter with internal 5 kV probe (HV
terminal)
10KV – multimeter with external 10 kV probe
40KV – multimeter with external 40 kV probe
Set 10kV external high voltage probe in
multimeter mode.
The Product returns OFF, 5KV, 10KV or
40KV.