[SOUR]:SAF:RCDC:CURR?
Description: Returns the measured trip current. The Product returns measured
trip current value using exponential format. Example: 1 mA is
returned as 1.0000e-03. If not already selected, this command
switches the Product into the RCD Trip Current function mode.
Query:
SAF:RCDC:CURR?
[SOUR]:SAF:RCDC:RPOS(?) <DNPD>
Description: This command selects one of the 16 available resistances. Index of
resistance is from 0 (25 mΩ) to 15 (2 kΩ). If not already selected,
this command switches the Product into the RCD Trip Current
function mode.
Selected resistance is also applicable for the RCDC Time function.
Parameters: <DNPD>
Example:
SAF:RCDC:RPOS 2
Query:
SAF:RCDC:RPOS?
[SOUR]:SAF:RCDC[:CURR]:NOM(?) <DPND>
Description: This command sets the nominal value of trip current. If not already
selected, this command switches the Product into the RCD Trip
Current function mode. Nominal value of trip current is also
applicable for the RCDC Time mode.
Parameters: <DNPD>
Example:
SAF:RCDC:NOM 1
Query:
SAF:RCDC:NOM?
[SOUR]:SAF:RCDC[:CURR]:MAX?
Description: This command returns the maximum (peak) value of trip current.
Maximum value can be cleared using command SAF:RCDC:CLEar.
The Product returns the current value using exponential format.
Example: 1 mA is returned as 1.0000e-03. If not already selected,
this command switches the Product into the RCD Trip Current
function mode.
Query:
SAF:RCDC:MAX?
Multifunction Electrical Tester Calibrator
Parameter represents index of required
resistance. Acceptable range is 0 ... 15.
Series resistance index 2 (100 mΩ)
The Product returns the index of selected
resistance using exponential format.
Example: 4 is returned as 4.
Parameter represents the nominal value of
the trip current expressed in Amps. See the
Specifications (found online at
www.Flukecal.com) for the acceptable
ranges.
Nominal Trip Time 1 A.
The Product returns the nominal value of
trip time using exponential format. Example:
200 mA is returned as 200e-03.
Supported SCPI Commands
127
Need help?
Do you have a question about the 5322A and is the answer not in the manual?